Event Code Message Text and Description Details and Solution
2043
Wellness test failed because of
top cover is missing.
Install the bottom cover on the
bottom module of the library. Also
check the module interconnect
cabling and module power cabling. If
the controller module cannot detect
both a top and bottom cover the
robot will not move.
2044
Wellness test failed because of
drive power board has failed.
1. Ensure that the drive power
boards are fully seated in the
module. See Installing the New
Drive Power Board in Spectra
Stack DC-DC Converter
Replacement Instructions.
2. Power cycle the Library.
2045
Wellness test failed because of
move media test failed.
Check for obstructions in the
pathway of the robot such as a
cartridge sticking out. Verify module
alignment and frame alignment.
Check if the robotic assembly is stuck
in the lock mechanism. Move robotic
assembly apart from lock mechanism
and enable lock mechanism correctly.
2046
Wellness test failed because of
drive communication test
failed.
Remove and reseat the drive tray to
ensure that the drive is fully seated.
If the issue persists then reset the
drive. Check the drive dump for
more information. See Downloading
Drive Logs on page203.
2047
Wellness test failed because
the barcode scanning test
failed.
Verify that there is no obstruction in
front of the barcode scanning module
on the cartridge table located on the
robotic assembly. If the error persists
replace the robotic assembly. See the
Spectra Stack Spool Robotic Replacement
Instruction.
JANUARY 2023 235
SPECTRA STACK LIBRARY
USER GUIDE
CHAPTER 9 - EVENT CODES