Section 5: Using a Keysight 4284/4980A LCR Meter Model 4200A-SCS Prober and External Instrument Control
5-10 4200A-913-01 Rev. A December 2020
Returned values are placed in the Analyze spreadsheet.
0: OK.
-10030 (HP4284_NOT_IN_KCON): No Keysight 4284A or Keysight 4980 LCR is defined in your
system configuration.
-10031 (HP4284_MEAS_ERROR): A measurement error occurred.
-10090 (GPIB_ERROR_OCCURRED): A GPIB communications error occurred.
-10091 (GPIB_TIMEOUT): A timeout occurred during communications.
-10100 (INVAL_PARAM): An invalid input parameter is specified.
-10102 (ERROR_PARSING): There was an error parsing the response from the Model 4284A or
4980.
-10101 (ARRAY_SIZE_TOO_SMALL): The specified value for Csize, G_or_Rsize, Vsize, or
Tsize was too small for the number of steps in the sweep.