EasyManuals Logo
Home>Tektronix>Measuring Instruments>Keithley 4200A-SCS

Tektronix Keithley 4200A-SCS User Manual

Tektronix Keithley 4200A-SCS
324 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #108 background imageLoading...
Page #108 background image
Section 5: Using a Keysight 4284/4980A LCR Meter Model 4200A-SCS Prober and External Instrument Control
5-12 4200A-913-01 Rev. A December 2020
Returned values are placed in the Analyze spreadsheet.
• 0: OK.
• -10000 (INVAL_INST_ID): The specified instrument ID does not exist.
• -10030 (HP4284_NOT_IN_KCON): No Keysight 4284A or Keysight 4980 LCR is defined in your
system configuration.
• -10031 (HP4284_MEAS_ERROR): A measurement error occurred.
• -10090 (GPIB_ERROR_OCCURRED): A GPIB communications error occurred.
• -10091 (GPIB_TIMEOUT): A timeout occurred during communications.
• -10102 (ERROR_PARSING): There was an error parsing the response from the Model 4284A
or 4980.