Model 4200A-SCS Prober and External Instrument Control Section 9: Using a Cascade Microtech PA200 Prober
4200A-913-01 Rev. A December 2020 9-9
Set up GPIB communications
To set up GPIB communications:
1. Double-click the ProberBench NT icon (shortcut) on desktop.
2. Double-click the Service Programs file.
3. Double-click the Prober Setup file in the Service Programs directory. The Prober Setup window
appears (see the following figure).
Figure 136: Prober setup: PC Config tab
4. Select the NI-GPIB option.
5. Clear the Simulate Prober box in the Server Mode section of the dialog box.
6. Click Save Setup.
7. From the ProberBench NT window, double-click the PB-GPIB file.