Section 4: Using a Model 590 C-V Analyzer Model 4200A-SCS Prober and External Instrument Control
4-22 4200A-913-01 Rev. A December 2020
InputPin
The DUT pin to which the 590 input terminal is attached (−1 to 72); if a value of less
than 1 is specified, no switch matrix connection is made; see Details
OutPin
The DUT pin to which the 590 output terminal is attached (−1 to 72); if a value of
less than 1 is specified, no switch matrix connection is made; see Details
OffsetCorrect
Determines if an offset correction measurement should be made:
â–ª Do not make offset measurement: 0
â–ª Make offset measurement: 1
FirstBias
The starting bias for the sweep: −20 V to +20 V; see Details
LastBias
The last voltage used in the sweep: −20 V to +20 V; see Details
StepV
The voltage step size: −20 V to +20 V; see Details
Frequency
The measurement frequency to use:
â–ª 100 kHz: 0
â–ª 1 MHz: 1
DefaultBias
The DC bias applied before and after a sweep: −20 V to +20 V
StartTime
The time that occurs on the first bias step, from the point the instrument is first
triggered until the first step time: 0.001 s to 65 s
StopTime
The time between pulses with the 590 at the default bias: 0.001 s to 65 s
StepTime
The time after a transition to a new bias step and before the instrument begins a
measurement: 0.001 s to 65 s
Range
The measurement range to use in F: 2E-12, 20E-12, 200E-12, or 2E-9; see Details
Model
Measurement model:
â–ª Series model: 0
â–ª Parallel model: 1
Filter
Enable or disable the analog filter, which can minimize the amount of noise that
appears in the readings; however, it increases the measurement time:
â–ª Disable the filter: 0
â–ª Enable the filter: 1
ReadingRate
Selects the reading rate used to acquire the measurements: 0 to 4; see Details
C
Output: The measured array of capacitance values
Csize
Set to a value that is equal to or greater than the G_or_Rsize or number of voltage
steps in the sweep, or is equal to
((LastBias − FirstBias) /StepV) + 1; when this function is called from a
Clarius, this parameter is fixed at 1350
V
The array of bias voltages used
Vsize
Set to a value that is equal to or greater than the G_or_Rsize or number of voltage
steps in the sweep, or is equal to
((LastBias − FirstBias) /StepV) + 1; when this function is called from a
Clarius, this parameter is fixed at 1350
G_or_R
Output:
â–ª When the parallel measurement model (1) is selected, G_or_R is the
measured conductance
â–ª When the series measurement model (0) is selected, this is the measured
resistance
G_or_Rsize
Set to a value that is equal to or greater than the G_or_Rsize or number of voltage
steps in the sweep, or is equal to
((LastBias − FirstBias) /StepV) + 1; when this function is called from a
Clarius UTM, this parameter is fixed at 1350

Table of Contents

Other manuals for Tektronix Keithley 4200A-SCS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 4200A-SCS and is the answer not in the manual?

Tektronix Keithley 4200A-SCS Specifications

General IconGeneral
BrandTektronix
ModelKeithley 4200A-SCS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals