4-2 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
System Specifications Chapter 4
Power Specifications CELL-DYN 3000SL
Table 4.3: Power Specifications
Consumption
Analyzer: 900 Watts
Data Station: 300 Watts
Printer: 145 Watts
Sample Loader: 50 Watts
1395 Watts maximum (4760 BTU per hour)
Analyzer Input Requirements
Setting Range Frequency
100 90—110 VAC 50/60 Hz
120 110—130 VAC 50/60 Hz
220 200—240 VAC 50/60 Hz
240 220—250 VAC 50/60 Hz
Data Station Input Requirements
Setting Range Frequency
120 90—130 VAC 50/60 Hz
240 180—260 VAC 50/60 Hz
Printer Input Requirements
Setting Frequency
120 50/60 Hz
Sample Loader Input Requirements
Setting Range Frequency
115 90—130 VAC 60 Hz
230 190—240 VAC 50 Hz