10-74 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting Chapter 10
Table 10.4: Non-Functional Fault Conditions
Analyzer or Data Station will not power ON
–––––Probable Cause––– ––––Corrective Action–––
1. Power source is defective. 1. Verify that the power
switch is OFF and
connect the system to a
known-good power source.
2. Power cord is not securely 2. Ensure that the power cord
connected to the Analyzer is securely connected to
or is not connected to the the Analyzer or Data
power outlet. Station and verify that it is
connected to the power
outlet.
3. Analyzer fuse is blown 3. The Analyzer fuse is located
or incorrect. above the power cord con-
nector on the rear panel.
Check the fuse as directed
in the Troubleshooting
Procedures section of this
chapter.
CAUTION: Always turn the Analyzer power switch OFF and
disconnect the power cord from the receptacle before checking
or replacing any fuse.
4. Defective power switch or 4. Call for technical assistance.
other system malfunction.
__________________________________________________________
No screen display
–––––Probable Cause––– ––––Corrective Action–––
1. Data Station power 1. Turn the power switch ON.
switch is turned OFF.
2. Data Station brightness 2. Adjust the brightness control
control is turned down. on the right side of the Data
Station until the image is
visible.
3. Defective Data Station 3. Call for technical assistance.
or other component.