10-62 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting Chapter 10
Table 10.2 General Fault Conditions (Continued)
Message acknowledgment time out/BL
–––––Probable Cause–––– ––––Corrective Action–––
1. Communication between 1. Initialize the Analyzer by
the Analyzer and the Data pressing the
Station did not occur when [INITIALIZATION] key on
expected. the second DIAGNOSTICS
MENU Screen. If the fault
recurs, call for technical
assistance.
__________________________________________________________
Message reception time out/BL
–––––Probable Cause–––– ––––Corrective Action–––
1. Communication between 1. Initialize the Analyzer by
the Analyzer and the Data pressing the
Station did not occur when [INITIALIZATION] key on
expected. the second DIAGNOSTICS
MENU Screen. If the fault
recurs, call for technical
assistance.
__________________________________________________________
Run time error/BL
–––––Probable Cause–––– ––––Corrective Action–––
1. An illegal software 1. Document what was
operation was requested happening when the
by the Analyzer. message was displayed and
call for technical assistance.
2. Initialize the Analyzer by
pressing the
[INITIALIZATION] key on
the second DIAGNOSTICS
MENU Screen. Resume
processing if the fault does
not recur.