EasyManua.ls Logo

Abbott CELL-DYN 3000 - Page 490

Abbott CELL-DYN 3000
542 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CELL-DYN 3000 System Operator’s Manual 10-61
9140240E — May 1995
Chapter 10 Troubleshooting
Table 10.2 General Fault Conditions (Continued)
Data acquisition overlap/BL
–––––Probable Cause–––– ––––Corrective Action–––
1. The timing of commun- 1. Document what was
ication between the happening when the
Analyzer and the Data message was displayed and
Station is incorrect. report the problem to the
Customer Support Center.
Initialize the Analyzer by
pressing the
[INITIALIZATION] key on
the second DIAGNOSTICS
MENU Screen. Resume
processing if the fault does
not recur.
NOTE: The error may occur when several tasks are requested in
rapid sequence. For example, print data log, transmit result,
sample processing, print ticket, print report, etc.
__________________________________________________________
List mode data phase error/BL
–––––Probable Cause––– ––––Corrective Action–––
1. The order of the data 1. Call for technical assistance.
received by the Data
Station during a measure-
ment was incorrect.
Search
Go Back
Book TOC

Table of Contents

Other manuals for Abbott CELL-DYN 3000

Related product manuals