CELL-DYN 3000 System Operator’s Manual 5-77
9140240E — May 1995
Chapter 5 Operating Instructions
15.Repeat steps 12–14 for all levels of controls that were run.
16.If the results are unacceptable, repeat the run. If the results are still
unacceptable, obtain a new tube of control, be sure that it is
warmed and mixed properly and again repeat the run. If the results
are still unacceptable, run the other levels of control material. If the
results on all levels are unacceptable, troubleshoot accordingly as
directed in the Troubleshooting Guide.
17.When the control results are acceptable, patient samples may be
analyzed.
Running Samples
Open Mode Analysis
The Open Sampler Mode aspirates the sample from an opened collection
tube. OPEN SAMPLER is displayed in the upper right corner of the
Data Station RUN Screen when this mode is selected. The open sample
aspiration probe is available only when this mode is selected.
Open Mode Procedure
1.If necessary, from the RUN Screen press [CHANGE SAMPLER]
to select the Open mode.
2.Be sure that the word “Ready” is illuminated on the Analyzer status
indicator panel and READY is displayed in the Status Box on the
Data Station RUN Screen.
WARNING: Potential Biohazard. Follow established
biosafety practices when performing maintenance, service or
troubleshooting procedures. Wear gloves, lab coat, and suitable
eye protection. Refer to Chapter 8, Precautions, Limitations and
Hazards, for additional information.
3.Open the sample tube and immerse the open sample aspiration
probe in the well-mixed sample.
4.Press the touch plate located behind the probe to start the cycle.
The word “Busy” on the Analyzer status indicator panel is
illuminated in yellow. The Status Box on the RUN Screen displays
messages to indicate the various stages of the cycle.
5.Remove the tube when the beep sounds. The probe moves up
through the wash block to clean it.