10-76 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting Chapter 10
Table 10.4 Non-Functional Fault Conditions (Continued)
The word FAULT on the Analyzer status indicator panel is
illuminated in red
NOTE: This type of fault is usually accompanied by a message in the
status box and/or on the bulletin line.
–––––Probable Cause––– ––––Corrective Action–––
1. The Analyzer has detected 1. From the first
a fault situation and has DIAGNOSTICS MENU
inhibited operation. Screen, press [FAULT
REPORT]. Print a
copy of the report and
perform the indicated
corrective action. When the
action is completed,
initialize the Analyzer by
pressing the
[INITIALIZATION] key on
the second DIAGNOSTICS
MENU Screen.
2. If the fault report does
not indicate a message or
action, document the
situation and initialize
the Analyzer by pressing the
[INITIALIZATION] key on
the second DIAGNOSTICS
MENU Screen.
__________________________________________________________
Instrument will not stop cycling
–––––Probable Cause––– ––––Corrective Action–––
1. The touch plate is stuck 1. Check the touch plate and
or being held down in remove any obstructions.
some way. Verify that it is not sticking by
pressing it several times.
2. Circuitry malfunction. 2. Call for technical assistance.