6-52 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Calibration Chapter 6
9. Press [PRINT QC LOG] to obtain a printout of the data.
10. Press [RETURN] followed by [MAIN] to return to the MAIN
MENU Screen.
Calibration Factor Calculations
1. Use the mean value for each parameter from the CD MEAN file
printout and the Current Open Mode Calibration Factor printout
from step 3 (Preparing for Manual Calibration section) for the
calibration factor calculations.
2. Enter the information from step 1 on the Manual Calibration
worksheet provided at the end of this procedure to calculate the
New Open Mode Calibration Factor for each parameter as follows.
CalibratorCalibration:
Whole Blood Calibration:
3. Referring to the New Open Mode Cal Factor from the previous step
and the Current Open Mode Cal Factor from the printout, use the
appropriate chart on the worksheet to compute the Factor %
Difference as follows:
New Open Mode Factor - Current Open Mode Factor x 100
New Open Mode Factor
= Factor % Difference
4. Enter the factor % difference for each parameter in the Manual
Calibration Criteria chart on the worksheet. The Calibration
Criteria are shown in Table6.5.
NOTE: Delete the sign of the factor % difference value before
entering it on the chart.
5. To determine which parameters require calibration, continue with
the next section.
Assay Value
CELL-DYN Mea
n
---------------------------------------------
Current Open Mode
Calibration Facto
r
×
New Open Mode
Calibration Facto
r
=
Reference Mean
CELL-DYN Mea
n
---------------------------------------------
Current Open Mode
Calibration Facto
r
×
New Open Mode
Calibration Facto
r
=