CELL-DYN 3000 System Operator’s Manual 6-67
9140240E — May 1995
Chapter 6 Calibration
5. Type any comments in the <COMMENTS> field. Press the Enter
key on the keyboard to save the entry and advance the cursor.
NOTE: Comments may be added to the Calibration Log only
after a calibration factor is changed or reentered.
6. If desired, press [PRINT LOG] to obtain a printout of the
Calibration Log for documentation.
Calibration for Individual Parameters
If only certain individual parameters require calibration, continue with
this section:
1. Press [QUIT AUTO-CAL] followed by [CONFIRM QUIT] to
exit Auto-Cal.
2. Press [RETURN] twice to return to the CALIBRATION MENU
Screen.
3. Press [ENTER FACTOR]
The ENTER WHOLE BLOOD FACTOR Screen is displayed.
4. Type any mean factors computed by Auto-Cal for those parameters
that require calibration in the appropriate fields. (Refer to the printout
made in step 2 of the “Calibration Factor Calculation” section of
this procedure.) Move the cursor to the desired position, type the
factor and press the Enter key on the keyboard to save the factor and
advance the cursor.
5. Press [RETURN].
6. When the CALIBRATION LOG Screen is displayed, indicate the
parameters that required calibration in the <COMMENTS> line.
Press the Enter key on the keyboard to save the entry and advance
the cursor.
NOTE: Manually entered factors are followed by the letter E in
parentheses to indicate “Enter Factor”.
7. If desired, press [PRINT LOG] to obtain a printout of the
Calibration Log for documentation.
Completing Mode to Mode Calibration
1. Press [RETURN] to return to the CALIBRATION MENU
Screen.
2. Press [MAIN] to exit the CALIBRATION MENU.
3. Retain all necessary documentation in the instrument logbook.