CELL-DYN 3000 System Operator’s Manual 6-17
9140240E — May 1995
Chapter 6 Calibration
Figure 6.4: Calibration Log Screen
Figure 6.5: Auto-Calibration Screen for CELL-DYN 3000
CLOSED
SAMPLER
PRINT
LOG
RETURN
CALIBRATION LOG
Ready
Aug 11 1992
Operator ID
Sequence #
10:35
734
4100
Open Sampler
Date Time OpID WBC RBC HGB MCV PLT MPV
04/28/92 14:45 734 1.010(A) 1.070(A) 1.120(A) 0.990(A) 1.030(A) 1.000(F)
Comments: CALIBRATION TO CALIBRATOR-LOT CD3012. INSTRUMENT INSTALLATION
08/02/92 11:41 734 1.010(A) 1.130(E) 1.120(A) 1.010(E) 1.130(E) 1.000(F)
Comments: RBC/PLT APERTURE REPLACED.
WHOLE
BLOOD
CALIBRATR LATEX CHANGE
SAMPLER
RETURN
AUTO-CALIBRATION
Ready
May 11 1992
Operator ID
Sequence #
10:30
734
4100
Open Sampler
Open sampler is selected. To calibrate using the
closed sampler, first press the CHANGE SAMPLER key.
To proceed, press the key for the specimen type being
used (WHOLE BLOOD, CALIBRATOR, or LATEX).