CELL-DYN 3000 System Operator’s Manual 8-1
9140240E — May 1995
Chapter 8
Precautions, Limitations and Hazards
Limitations
The CELL-DYN 3000® is designed for in vitro diagnostic use.
Abbott has designed the CELL-DYN 3000 system components for
optimal performance. Substitution of reagents, calibrators, controls
and components manufactured by other companies may adversely
affect the performance of the Analyzer.
Follow the recommended maintenance schedules and procedures as
outlined in Chapter 9, Maintenance, of this manual.
All service and repair must be performed by authorized Abbott
representatives.
Location Requirements
An authorized Abbott representative must install the instrument.
Place the CELL-DYN 3000 Analyzer and Data Station on a hard,
level surface. The location should have nonporous, non-absorbing
work surfaces and flooring which can be easily cleaned and
disinfected using recommended procedures. Locate the system:
Away from direct sunlight
Away from the path of a cooled or heated air outlet
Away from centrifuges, x-ray equipment, CRTs, video
terminals, computers and copiers
Do not place reagent containers above the instrument.