CELL-DYN 3000 System Operator’s Manual 8-1
9140240E — May 1995
Chapter 8
Precautions, Limitations and Hazards
Limitations
• The CELL-DYN 3000® is designed for in vitro diagnostic use.
• Abbott has designed the CELL-DYN 3000 system components for
optimal performance. Substitution of reagents, calibrators, controls
and components manufactured by other companies may adversely
affect the performance of the Analyzer.
• Follow the recommended maintenance schedules and procedures as
outlined in Chapter 9, Maintenance, of this manual.
• All service and repair must be performed by authorized Abbott
representatives.
Location Requirements
• An authorized Abbott representative must install the instrument.
• Place the CELL-DYN 3000 Analyzer and Data Station on a hard,
level surface. The location should have nonporous, non-absorbing
work surfaces and flooring which can be easily cleaned and
disinfected using recommended procedures. Locate the system:
• Away from direct sunlight
• Away from the path of a cooled or heated air outlet
• Away from centrifuges, x-ray equipment, CRTs, video
terminals, computers and copiers
• Do not place reagent containers above the instrument.