10-54 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting Chapter 10
Table 10.2: General Fault Conditions
The following abbreviations are used in this table:
BL — Bulletin Line
SB — Status Box
Information for each message is listed in order from the most likely
cause of the message to the least likely cause of the message. Therefore,
always troubleshoot a problem in this order. If a problem cannot be
resolved by the corrective action indicated in this table, call for technical
assistance.
__________________________________________________________
Not Ready: See DIAG or Not Ready: See Special/SB
If a fault condition has occurred, the word FAULT on the Analyzer status
indicator panel is illuminated in red.
––––Probable Cause–––– ––––Corrective Action––––
1. A situation that prevents 1.From the first DIAGNOSTICS
the READY state has been MENU Screen, press
detected. See the [FAULT REPORT] followed
DIAGNOSTICS MENU by [PRINT] to obtain a
Screen or the SPECIAL printout describing the
PROTOCOLS MENU Screen, problem. Refer to the
whichever is indicated, for for appropriate table for more
information. corrective action.
2. A diagnostic test was run 2.The instrument must be using
one of the initialized when the
DIAGNOSTICS MENU diagnostic test in progress
Screen keys. is complete. Initialize the
Analyzer by pressing the
[INITIALIZATION] key on
the second DIAGNOSTICS
MENU Screen.
3. A special protocols 3.Check the SPECIAL
procedure is in progress. PROTOCOLS MENU Screen
and perform the action
necessary to complete the
procedure.
4. System malfunction. 4.Call for technical assistance.