Chapter 10 Troubleshooting
Table 10.2 General Fault Conditions (Continued)
Uninitialized/SB
––––Probable Cause–––– ––––Corrective Action––––
1. Data Station has power 1. Ensure that the Analyzer but
the Analyzer is not power cord is connected to
responding. the Analyzer and that the
cord is connected to the
power outlet. Initialize the
Analyzer by pressing the
[INITIALIZATION] key on
the second DIAGNOSTICS
MENU Screen.
2. Communication mal- 2. Check the cable that function
between the connects the Analyzer
Analyzer and Data Station. to the Data Station. If
necessary, secure the
connections. Initialize the
Analyzer by pressing the
[INITIALIZATION] key on
the second DIAGNOSTICS
MENU Screen.
__________________________________________________________
Initialization Failed
Bottom of screen (MAIN MENU is not displayed)
–––––Probable Cause––––– –––– Corrective Action––––
1. The Data Station was 1. Initialize the Analyzer by
unable to initialize. The following the instructions
CELL-DYN software does in the Power OFF and
not display the Power ON procedures
MAIN MENU Screen. given in the Troubleshooting
Procedures section of this
chapter. If the Data Station
does not initialize, press
the Print Screen key on the
computer keyboard to
document any screen
messages, then call for
technical assistance.