EasyManua.ls Logo

Abbott CELL-DYN 3000 - Power Specifications - CELL-DYN 3000 CS; Consumption

Abbott CELL-DYN 3000
542 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CELL-DYN 3000 System Operator’s Manual 4-7
9140240E — May 1995
Chapter 4 System Specifications
Power Specifications CELL-DYN 3000CS
Table 4.6: Power Specifications
Consumption
Analyzer: 900 Watts
Data Station: 300 Watts
Printer: 145 Watts
1345 Watts maximum (4590 BTU per hour)
Analyzer Input Requirements
Setting Range Frequency
100 90—110 VAC 50/60 Hz
120 110—130 VAC 50/60 Hz
220 200—240 VAC 50/60 Hz
240 220—250 VAC 50/60 Hz
Data Station Input Requirements
Setting Range Frequency
120 90—130 VAC 50/60 Hz
240 180—260 VAC 50/60 Hz
Printer Input Requirements
Setting Frequency
120 VAC 50/60 Hz
Search
Go Back
Book TOC

Table of Contents

Other manuals for Abbott CELL-DYN 3000

Related product manuals