3-4 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Principles of Operation Chapter 3
Instrument Rinsed
1. The open sample aspiration probe is rinsed internally and
externally with Diluent.
2. The needle used in either closed mode is rinsed internally and
externally with Diluent.
3. The WBC mixing chamber and WBC flow cell are rinsed with
Sheath.
4. The RBC/PLT transducer assembly is rinsed with Diluent.
5. The RBC/PLT metering tube is coated with Hemoglobin Lyse.
6. The HGB mixing chamber and HGB flow cell are rinsed with
Hemoglobin Lyse.