1-22 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
System Description Chapter 1
Printer
The printer is discussed in detail in Chapter 11, Printer.
Sample Loader
The Sample Loader is discussed in detail in Chapter 12, Sample Loader.
Reagent System
Introduction
The reagent system is formulated specifically for the CELL-DYN 3000
instrument flow systems in order to provide optimal system performance.
Use of reagents other than those specified in this manual is not
recommended as instrument performance can be affected. Each
CELL-DYN 3000 system is tested at the factory using the specified
reagents and all performance claims were generated using these reagents.
Reagents must be stored at room temperature to ensure optimal
performance. All reagents should be protected from direct sunlight,
extreme heat and freezing during storage. Temperatures below 32°F
(0°C) may cause reagent layering that changes the tonicity and
conductivity of the reagents. If any reagent has been frozen, it should not
be used.
The reagent inlet tubes have a cap attached that minimizes evaporation
and contamination during use. To ensure optimal performance and
accurate results, use all reagents within the dating period indicated on the
label.