6-14 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Calibration Chapter 6
Figure 6.2: Calibration Menu Screen Displaying Closed Mode
Calibration Factors
The function of each key is discussed briefly in this section. The Auto-
Cal Calibration Procedures provide detailed instructions for using the
Auto-Cal program.
NOTE: For ease of explanation, the key labels may not always
be discussed in the order in which they appear on the screen.
The [OPEN SAMPLER/CLOSED SAMPLER] key is used to display
the current calibration factors for the mode selected (see Figures6.1
and6.2).
NOTE: The key is labeled for the sampler mode that is NOT
currently displayed.
The [PRINT] key is used to print the Current Whole Blood Factors
displayed on the CALIBRATION MENU Screen.
ENTER
FACTOR
CALIBRATN
LOG
AUTO-
CALIBRATE
OPEN
SAMPLER
PRINT MAIN
CALIBRATION
Ready
May 11 1992
Operator ID
Sequence #
10:30
734
4100
Closed Sampler
Current Whole Blood Factors:
Parameter Method Factor Date Time Operator
WBC WHOLE BLOOD 1.056 04/30/92 16:59 C03
RBC WHOLE BLOOD 1.060 04/30/92 16:59 C03
HGB WHOLE BLOOD 1.136 04/30/92 16:59 C03
MCV WHOLE BLOOD 0.985 04/30/92 16:59 C03
PLT WHOLE BLOOD 1.039 04/30/92 16:59 C03
MPV FACTORY 1.000 03/02/92 11:41 C03