CELL-DYN 3000 System Operator’s Manual 4-11
9140240E — May 1995
Chapter 4 System Specifications
WBC Differential Parameters
Precision specifications for the WBC Differential parameters are given as
a 95% confidence limit for a range of values for each of the WBC
subpopulations.
Table 4.8: Precision of the WBC Differential Parameters
Linearity
Linearity specifications are determined by analyzing dilutions of a
commercially available control material that contains no interfering
substances and displays no suspect parameter flags. Specifications are
determined by taking multiple measurements on each dilution to
minimize the effect of imprecision. The stated limits are determined by
regression analysis with the line going through the origin (0,0).
Table 4.9: Linearity Specifications
Cell Type Range 95% Confidence Limit
Neutrophil % 42—65% ± 2.1%
Lymphocyte % 24—45% ± 3.2%
Monocyte % 3.6—11.0% ± 2.9%
Eosinophil % 1.0—9.5% ± 1.2%
Basophil % 0.5—1.3% ± 1.0%
Parameter Linear Range
Acceptable Limits
(whichever is greater)
WBC 0.0—99.9 K/µL ± 0.4 or 3.0%
RBC
0.00—7.00 M/µL
± 0.1 or 2.5%
HGB 2.5—24 g/dL ± 0.3 or 2.0%
MCV 50—200 fL ± 3.0 or 3.0%
PLT 0—999 K/µL ± 12.0 or 4.0%
MPV 5—20.0 fL ± 1.0 or 6.0%