CELL-DYN 3000 System Operator’s Manual 5-21
9140240E — May 1995
Chapter 5 Operating Instructions
Figure 5.10: Replicate ID Entry Screen
2.<LOT NUMBER> and <EXPIRATION DATE> entry fields (see
the following figure)
These entry fields are displayed when the [LOT NUMBER] key is
pressed. This designation is intended for QC Files that are used for
commercial controls.
3.WESTGARD RULE SELECTION:
RULE 1: Value Outside 3 SD
RULE 2: Two Consecutive values outside SAME 2 SD
RULE 3: Two Consecutive values outside OPPOSITE 2 SD
RULE 4: Two of Three Consecutive values outside SAME 2 SD
RULE 5: Four Consecutive values outside SAME 1 SD
RULE 6: Ten Consecutive values on SAME side of mean.
NOTE: Westgard Rules are discussed in detail in the Quality
Control chapter of this manual.
The Westgard Rule selections are available on either of the QC FILE
SETUP Screens.
LOT
NUMBER
PRINT RETURN
QC FILE SETUP
Ready
Dec 17 1992
Operator ID
Sequence #
16:36
734
0630
Replicate ID: ------------
FOR Low
WESTGARD RULE SELECTION:
ON
ON
ON
ON
ON
ON
RULE 1:
RULE 2:
RULE 3:
RULE 4:
RULE 5:
RULE 6:
Value outside 3 SD.
Two consecutive values outside SAME 2 SD.
Two consecutive values outside OPPOSITE 2 SD.
Two of three consecutive values outside SAME 2 SD.
Four consecutive values outside SAME 1 SD.
Ten consecutive values on SAME side of mean.
1