CELL-DYN 3000 System Operator’s Manual 5-83
9140240E — May 1995
Chapter 5 Operating Instructions
NOTE: The touch plate will not operate if the tube is not seated
correctly. (If necessary, refer to the Special Procedures section in
Chapter 9, Maintenance, for instructions for adjustment of the
Tube Retainer.)
4.Press the touch plate located behind the probe to start the cycle.
The word “Busy” on the Analyzer status indicator panel is
illuminated in yellow.
5.Remove the tube when the beep sounds.
6.When the cycle is completed, the word “Ready” on the Analyzer
status indicator panel is illuminated in green and the results are
displayed on the Data Station RUN Screen.
7.Repeat this procedure for subsequent samples.
Daily Shutdown
It is not necessary to perform a Daily Shutdown procedure as the
instrument automatically goes into a STANDBY state if it has been idle
for four hours. If desired, the operator may place the instrument in the
STANDBY state by pressing [SPECIAL PROTOCOLS], followed by
[MORE], followed by [DAILY SHUTDOWN].
When the key is pressed or when the automatic shutdown is initiated, the
cycle:
Rinses the flow system
Sets the timer control that periodically opens all of the solenoid
valves to prevent pinched tubing