CELL-DYN 3000 System Operator’s Manual 5-73
9140240E — May 1995
Chapter 5 Operating Instructions
Sample Analysis on the CELL-DYN 3000SL
Daily Start Up Procedure
The automatic Start Up cycle is designed to prime the flow system and
run background counts whenever the STANDBY or INITIALIZED
message appears in the Status Box on the Data Station RUN Screen. The
cycle takes approximately 3.5 minutes and is activated by pressing the
[RUN] or [PRIME] key. (The [PRIME] key will be displayed instead of
the [CLEAR ORIFICE] key if the instrument goes into STANDBY
while in the RUN screen.)
Before beginning this procedure, ensure that power to all components is ON.
The power to the Analyzer should always be ON.
Start Up Procedure
1.Be sure that the word “Ready” on the Analyzer status indicator
panel is illuminated in green and the READY message is displayed
in the Status Box on the Data Station RUN Screen.
2.If the Status Box on the Data Station RUN Screen displays
STANDBY or INITIALIZED, press [RUN] or [PRIME] to initiate
the automatic Start Up cycle.
3.Be sure that all 10 racks are in the sample loader tray (5 on each
side) and the safety cover is in place.
4.Turn the Sample Loader power switch ON. When the Sample
Loader initialization cycle is completed, the indicator on the
Sample Loader Start key blinks and the Bulletin line on the Data
Station RUN Screen displays the message
AUTO-SAMPLER READY.
5.When the automatic start up cycle is completed, a background
count is automatically performed and the Open mode is selected.
6.Verify that the background counts are acceptable. If the
background counts are unacceptable, repeat the background cycle.
If the counts are still unacceptable, troubleshoot accordingly as
directed in the Troubleshooting Guide.
NOTE: Backgrounds may be repeated by pressing the touch
plate.
7.When the background counts are acceptable, perform the daily
Quality Control checks as directed in the following section.