CELL-DYN 3000 System Operator’s Manual 10-75
9140240E — May 1995
Chapter 10 Troubleshooting
Table 10.4 Non-Functional Fault Conditions (Continued)
The MAIN MENU Screen is not displayed after initialization
–––––Probable Cause–––– –––– Corrective Action–––
1. There is a floppy disk in 1. If a disk is present, remove
the Data Station disk drive. it and initialize the Data
Station by turning the
power switch OFF and ON.
2. The external keyboard 2. Disconnect the external
is defective. keyboard from the Data
Station and then initialize
the Data Station by turning
the power switch OFF and
ON. If the MAIN MENU
Screen is displayed after
initialization is complete,
call the Customer Support
Center and report the
problem. Resume operation
using the Data Station
membrane keyboard.
3. A file is missing or is 3. Call for technical assistance.
incorrect on the Data
Station’s hard drive.
4. There is a hardware or 4. Call for technical assistance.
software malfunction.