10-60 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting Chapter 10
Table 10.2 General Fault Conditions (Continued)
Flow sequence time out <x,x>/BL
NOTE: Characters in the brackets identify the problem flow
sequence number and flow sequence name.
–––––Probable Cause–––– ––––Corrective Action–––
1. An internal Analyzer flow 1. Record the characters in the
sequence was not completed brackets. Call the Customer
in the allotted time. Support Center to report
the problem at
1 (800) CELL DYN.
Initialize the Analyzer by
pressing the
[INITIALIZATION] key on
the second DIAGNOSTICS
MENU Screen. Resume
processing if the fault does
not recur.
__________________________________________________________
Command sent at incorrect time/BL
–––––Probable Cause–––– ––––Corrective Action–––
1. The Analyzer received a 1. Initialize the Analyzer by
command from the Data pressing the
Station at the incorrect [INITIALIZATION] key on
time and could not the second DIAGNOSTICS
process it. MENU Screen. If necessary,
call for technical assistance.