CELL-DYN 3000 System Operator’s Manual 10-63
9140240E — May 1995
Chapter 10 Troubleshooting
Table 10.2 General Fault Conditions (Continued)
Bad checksum in non-volatile RAM/BL
–––––Probable Cause–––– ––––Corrective Action––––
1. When the system was 1. Call for technical assistance.
powered ON, the Analyzer
did not transmit the correct
message to the Data Station.
__________________________________________________________
Bad monitor command/BL
NOTE: This message does not pertain to the Data Station
screen (CRT).
–––––Probable Cause–––– ––––Corrective Action––––
1. During the initialization 1. Initialize the Analyzer
process, the Data Station by pressing the
and Analyzer did not [INITIALIZATION] key
communicate properly. on the second
DIAGNOSTICS MENU
Screen. If the fault recurs,
call for technical assistance.