CELL-DYN 3000 System Operator’s Manual 5-17
9140240E — May 1995
Chapter 5 Operating Instructions
Figure 5.8: QC Range Entry Screen
Procedure Range Entry
1.Select a file from the QC SETUP MENU Screen by using the
Arrow keys on the keyboard to move the cursor into the desired
file.
2.Press [QC LIMITS] to display the QC RANGE ENTRY Screen
for the selected file.
3.Use the Arrow keys on the keyboard to move the cursor to the
desired entry field.
4.Type the numbers and press the Enter key on the keyboard to save
each entry and advance the cursor to the next entry field.
5.Repeat step 4 until all entries have been made.
6.If desired, press [PRINT] to obtain a printout of the entered values.
MEANS/
LIMITS
UPDATE
FROM FILE
LOAD
FROM DISK
PRINT RETURN
QC RANGE ENTRY
Ready
Jun 08 1992
Operator ID
Sequence #
10:23
734
9110
FOR NORM RD43 SL
WBC:
NEU:
LYM:
MONO:
EOS:
BASO:
RBC:
HGB:
HCT:
MCV:
MCH:
MCHC:
RDW:
PLT:
MPV:
PCT:
PDW:
Lower Limits
6.9
3.5
2.4
0.6
0.0
0.0
4.20
13.1
36.3
81.0
27.9
33.2
13.0
220.
6.0
0.21
14.5
K/uL
K/uL
K/uL
K/uL
K/uL
K/uL
M/uL
g/dL
%
fL
pg
g/dL
%
K/uL
fL
%
10(GSD)
47.3
31.5
7.0
1.2
0.0
%N
%L
%M
%E
%B
Upper Limits
8.5
4.3
3.0
1.0
0.4
0.2
4.50
13.9
40.3
89.0
31.9
37.2
17.0
270.
10.0
0.23
15.9
K/uL
K/uL
K/uL
K/uL
K/uL
K/uL
M/uL
g/dL
%
fL
pg
g/dL
%
K/uL
fL
%
10(GSD)
55.3
37.5
13.0
5.2
2.0
%N
%L
%M
%E
%B