CELL-DYN 3000 System Operator’s Manual 5-57
9140240E — May 1995
Chapter 5 Operating Instructions
The Spec ID and Sequence Number for the specimen displayed appears
in the upper left hand corner of the display and printout. The information
in the upper right hand corner displays the current operational status of
the Analyzer. This includes the current date, time, operator ID, current
specimen sequence number and sampler status (Open or Closed).
The data displayed on the left side of the FLAGGING DIAGNOSTICS
Screen and printout provide supplemental information regarding the
suspect population flags for BAND, IG, VAR LYM, VLYM/BLAST and
NRBC. The word SUSPECT will also appear above the suspect
population flags as it does on the RUN and DISPLAY SPECIMEN
Screens. The number of cells found in each region associated with the
suspect population flag is displayed as a percentage of the total WBC
value. The region percentages displayed on the FLAGGING
DIAGNOSTICS Screen are used as criteria for the suspect population
flags displayed on the RUN Screen and the DISPLAY SPECIMEN
Screen, thus providing additional information as to potential severity of the
flag.
The graphs located on the right side of the FLAGGING
DIAGNOSTICS Screen are determined by the parameter set which had
been selected for the specimen displayed. The parameter set is indicated
on the RUN Screen or DISPLAY SPECIMEN Screen. On the most
current 2000 specimens in the Data Log, the parameter set can be edited
after the sample has been processed using the [EDIT SPECIMEN] soft
key on the DATA LOG Screen.
NOTE: These data are provided for laboratory use only. The
information may be incorporated into your laboratory's review
criteria and guidelines. Whenever a morphological flag is
displayed, a stained smear should be reviewed.