5-90 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Operating Instructions Chapter 5
3.Limit Set entry selected
This field is used to specify which Patient Limits are assigned to
each sample. The Default Limit Set (see option 4) is used if no
specification is made.
4.Default Limit SET (1...4)
This field is used to specify the default (preassigned) Limit Set that
is automatically assigned to each sample unless otherwise indicated
in the Work List.
5.Parameter Set entry selected
This field is used to specify which Parameter Set is assigned to
each sample. The Default Parameter Set (see option 6) is used if no
specification is made.
6.Default Parameter Set (1...4)
This field is used to specify the default (preassigned) Parameter Set
that is automatically assigned to each sample unless otherwise
indicated in the Work List.
The [PRINT WORK LIST] key is used to print the Work List.
The [RETURN] key is used to return to the Data Station RUN Screen.
Work List Setup Procedure
1.From the Data Station RUN Screen, press [WORK LIST]
followed by [WORK LIST SETUP] to display the WORK LIST
SETUP Screen.
2.Select the type of bar code that will be used:
Type 1 to use the CELL-DYN Series 4-digit bar code
Type 2 to use a laboratory-generated bar code
Press the Enter key on the keyboard to save the selection and
advance the cursor.
PRINT
WORK LIST
RETURN