EasyManua.ls Logo

Abbott CELL-DYN 3000 - Page 305

Abbott CELL-DYN 3000
542 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-50 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Calibration Chapter 6
CELL-DYN 3000
Auto-Cal Calibration Worksheet
Instrument:______________________________________
Date:________________________________________
Operator:________________________________________
________________________________________________________________________________________
_______
* Delete the sign of the Factor % Difference before entering it on the chart.
** Do not calibrate. Call the Customer Support Center for assistance at 1 (800) CELL DYN.
Auto-Cal Calibration Criteria
Factor
%Diff*
Validation Range
Cal Not Required
Cal Range
Cal Required
Cal Limit
Do Not Cal**
Cal?
Y/N
WBC
<1.5% >1.5% but <10%
>10%
RBC
<1.0% >1.0% but <10%
>10%
HGB
<1.0% >1.0% but <10%
>10%
MCV
<1.0% >1.0% but <10%
>10%
PLT
<3.0% >3.0% but <15%
>15%
Search
Go Back
Book TOC

Table of Contents

Other manuals for Abbott CELL-DYN 3000

Related product manuals