6-50 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Calibration Chapter 6
CELL-DYN 3000
Auto-Cal Calibration Worksheet
Instrument:______________________________________
Date:________________________________________
Operator:________________________________________
________________________________________________________________________________________
_______
* Delete the sign of the Factor % Difference before entering it on the chart.
** Do not calibrate. Call the Customer Support Center for assistance at 1 (800) CELL DYN.
Auto-Cal Calibration Criteria
Factor
%Diff*
Validation Range
Cal Not Required
Cal Range
Cal Required
Cal Limit
Do Not Cal**
Cal?
Y/N
WBC
<1.5% >1.5% but <10%
>10%
RBC
<1.0% >1.0% but <10%
>10%
HGB
<1.0% >1.0% but <10%
>10%
MCV
<1.0% >1.0% but <10%
>10%
PLT
<3.0% >3.0% but <15%
>15%