9-10 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Maintenance Chapter 9
Figure 9.3: Special Protocols Screen 2
The [FLUSH SHEATH] key is used to flush the WBC flow cell with
Sheath reagent.
The [AUTO CLEAN] key is used to initiate the Auto-Clean procedure.
The shear valve, the WBC mixing chamber, the WBC flow cell, the HGB
mixing chamber, the HGB flow cell, the RBC/PLT transducer assembly
and all of the associated fluidics are automatically cleaned and rinsed
during the procedure.
When the [AUTO CLEAN] key is pressed, the screen displays the
message shown in Figure9.5, Special Protocols: Auto-Clean Screen.
The following soft key labels are displayed:
START
CANCEL
These keys are used to start or cancel the Auto-Clean procedure.
FLUSH
SHEATH
AUTO
CLEAN
DAILY
SHUTDOWN
PREPARE
SHIPPING
CLEAN
NEEDLE
EXTEND
AUTOCLEAN
MORE MAIN
SPECIAL PROTOCOLS
Ready
Mar 11 1993