9-14 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Maintenance Chapter 9
CAUTION: Use powder-free gloves when performing the
maintenance procedure to avoid contaminating the instrument.
Materials Required
1. CELL-DYN Enzymatic Cleaner (brought to room temperature)
2. Clean test tube or container
3. Lint-free wipes or gauze
4. Warm water
5. Gloves, lab coat, and suitable eye protection
Procedure
1. Select the open sampler mode.
2. Carefully wipe the outside of the open sampler probe and the
bottom of the wash block with a piece of gauze dampened with
warm water and a few drops of CELL-DYN Enzymatic Cleaner.
3. From the MAIN MENU Screen, Press [SPECIAL
PROTOCOLS] followed by [MORE] to access the Auto-Clean
function.
4. Press [AUTO CLEAN].
Instructions for performing the procedure are given on the screen
(See Figure9.5).
5. Dispense approximately 1.5 mL of enzymatic cleaner (undiluted)
into a clean container and hold the container under the open
sampler probe.
NOTE: Do not press the touch plate. The Auto Clean procedure
is only initiated by the [START] key.
6. Press [START].
7. Continue to hold the container under the probe until a beep tone is
heard. Remove the container and discard the remaining enzymatic
cleaner.
8. When the Auto-Clean procedure is completed, press [MAIN] to
return to the MAIN MENU Screen.