CELL-DYN 3000 System Operator’s Manual 9-23
9140240E — May 1995
Chapter 9 Maintenance
WARNING: Potential Biohazard. Follow established
biosafety practices when performing maintenance, service or
troubleshooting procedures. Wear gloves, lab coat, and suitable
eye protection. Refer to Chapter 8, Precautions, Limitations and
Hazards, for additional information.
CAUTION: Use powder-free gloves when performing the
maintenance procedure to avoid contaminating the instrument.
Procedure
1. Remove the racks from the Sample Loader tray.
2. Wash the racks in the detergent solution. Do not allow them to soak
in the solution, as the labels will come off. Use cotton swabs to
dislodge any debris.
3. Verify rack labels are intact and undamaged.
NOTE: Do not wash the racks in an automated dishwasher that
operates at high temperatures because the heat may damage the
racks.
4. Rinse the racks with warm water and dry thoroughly with lint-free
wipes or towels.
5. Wipe the stainless steel tray area with a lint-free wipe moistened
with water. Dry the tray with a lint-free wipe or towel.
6. Wipe the stainless steel plate and glass bar code reader window
beneath the tower with a lint-free wipe moistened with water. Dry
with a lint-free wipe.
7. Clean the mixing head with a lint-free wipe moistened with water.
Dry the head with a lint-free wipe.
8. Wash the safety cover with the detergent solution, rinse and dry it.