9-34 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Maintenance Chapter 9
7. Press [START].
8. Continue to hold the container under the probe until a beep tone is
heard. Remove the container and discard the remaining enzymatic
cleaner.
NOTE: The complete procedure takes 2.5 hours but may be
terminated after 38 minutes by pressing the [CANCEL] key. If
the [CANCEL] key is pressed, a rinse cycle is initiated which
prepares the instrument for sample processing. When the rinse
cycle finishes, press [MAIN] to return to the MAIN MENU
Screen.
9. At the end of the 2.5 hours, the instrument automatically goes into
the STANDBY mode. Press [RUN] to bring the Analyzer out of
STANDBY and prepare it for sample processing.
10. Verify that the background counts are acceptable by running 5
background counts before running controls. If background counts
or controls are unacceptable, refer to Chapter 10, Troubleshooting.