CELL-DYN 3000 System Operator’s Manual 10-29
9140240E — May 1995
Chapter 10 Troubleshooting
During the initialization process, the Data Station software is accessed
and downloaded to the Analyzer. Once this is accomplished, all Analyzer
motors and pumps are moved to their “home” positions. (Increased
motor noises are normal during this process.) When the initialization has
been successfully completed, the message INITIALIZED is displayed
in the status box.
The reagent priming step is necessary after any initialization. This is
accomplished by pressing [RUN] or [PRIME], whichever is displayed.
All reagents are primed automatically, a background cycle is performed
and the results are displayed on the RUN Screen. When the reagents are
primed, the message READY is displayed in the status box.
On the CELL-DYN 3000SL, the Sample Loader may be initialized by
pressing the INIT key on the Sample Loader operation keyboard. It may
also be initialized by turning the Sample Loader power switch OFF and
ON. (The Sample Loader is initialized every time the power switch is
turned ON.)