10-70 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting Chapter 10
Table 10.3 Sample-Related Fault Conditions (Continued)
Auto-Sampler consecutive data faults/BL
–––––Probable Cause––– ––––Corrective Action–––
1. During sample processing, 1. Correct the situation on the
four consecutive incomplete Analyzer as follows:
aspiration or metering faults
were detected and the
Check the appropriate
Sample Loader halted.
tubing for a plug
or pinch.
Press [CLEAR ORIFICE]
to clear the aperture.
Clean the Sample Loader
needle. (If necessary,
refer to the instructions
given in Chapter 9,
Maintenance.)
2. Press the [CLEAR FAULT]
key displayed on the RUN
Screen. The Sample Loader
automatically resumes
processing.
3. If the problem persists,
follow the instructions
given in the Trouble-
shooting Tips and
Techniques section of this
chapter for trouble-
shooting clog, flow or
incomplete aspiration
messages.