Repairing the Fluorescence Detector 5
1100 Series FD Reference Manual 157
5
Use a toothpick to remove the window. Note:
The quartz window may be cleaned with alcohol
(e.g. ethanol).
6
Insert the cleaned or new window. 7 Slide the window holder across the window and fix the
screws.