Repairing the Fluorescence Detector 5
1100 Series FD Reference Manual 169
2
Carefully remove the front foam holding the foam as
shown.
Note:
Locate the condenser assembly and the cutoff filter.
3
The cutoff filter can be replaced just by sliding it out and
back in.
4 Use a 2.5 mm hexagonal key to loosen the screw.