EasyManuals Logo

Agilent Technologies 1100 Series Reference Manual

Agilent Technologies 1100 Series
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #187 background imageLoading...
Page #187 background image
Repairing the Fluorescence Detector 5
1100 Series FD Reference Manual 189
6 Bring the top of the assembly into the position shown
below and fix the 3 screws.
7 Assure the UV protector cover is installed. Replace the
foam on the optical unit.
Next Steps:
Reconnect the encoder cable of the EX-monochromator to FLF board.
Reinstall the optical unit as described in Installing the Optical Unit" on page 194.
Enter the replacement in the maintenance logbook.
Reset the monochromator settings to default using the diagnose functions.
In this case “Wavelength Calibration Lostis displayed and the instrument stays in Not Ready condition until you
perform a wavelength calibration.
Perform a wavelength verification to check the correct positioning of the replaced assemblies, as described in chapter
“Wavelength Verification and Calibration" on page 101.

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals