EasyManua.ls Logo

Agilent Technologies 1100 Series - Page 190

Agilent Technologies 1100 Series
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
192 1100 Series FD Reference Manual
5 Repairing the Fluorescence Detector
5 Insert the power supply into its location and fix it with the
screws at the rear panel.
6 Press down and clip in the power switch light pipe into the
power supply.
7 Reinstall bottom foam piece. Slide it in underneath the
leak drain.
8 Reinstall the processor board, see “Exchanging the
Processor Board" on page 159.
9 Reinstall the fan assembly, see Exchanging the Fan" on
page 163.
10 Reinstall the optical unit, see Installing the Optical
Unit" on page 194.
11 Reinstall the front cover, top cover, see “Replacing the
Foams and Covers" on page 196.
12 Replace detector into the stack.
13 Reconnect the power cable and turn on the detector.