EasyManua.ls Logo

Agilent Technologies 1100 Series - Page 225

Agilent Technologies 1100 Series
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Identifying Parts and Materials 6
1100 Series FD Reference Manual 227
Agilent 1100 to HP 1050, HP 1046A or Agilent 35900 A/D Converters
Agilent 1100 to HP 1090 LC, HP 1040 DAD or Signal Distribution Module
Tabl e 46
Connector
5061-3378
Pin
HP 1050 /
Pin
Agilent 1100
Signal Name Active
(TTL)
1 - White 1 - White Digital ground
2 - Brown 2 - Brown Prepare run Low
3 - Gray 3 - Gray Start Low
4 - Blue 4 - Blue Shut down Low
5 - Pink 5 - Pink Not connected
6 - Yellow 6 - Yellow Power on High
7 - Red 7 - Red Ready High
8 - Green 8 - Green Stop Low
9 - Black 9 - Black Start request Low
Tabl e 47
Connector
01046-60202
Pin
HP 1090
Pin
Agilent 1100
Signal Name Active
(TTL)
1 1 - White Digital ground
NC 2 - Brown Prepare run Low
4 3 - Gray Start Low
74 - BlueShut downLow
8 5 - Pink Not connected
NC 6 - Yellow Power on High
3 7 - Red Ready High
6 8 - Green Stop Low
NC 9 - Black Start request Low
5 - Key