First Steps with the Fluorescence Detector 2
1100 Series FD Reference Manual 53
2 Wait until the baseline stabilizes. Complete the run.
3 Load the signal. (In this example just the time range of 13 minutes is
displayed).
Figure 18 Detector Settings for Emission Scan
DO NOT select
additional emission
wavelengths (B, C,
D). Doing so will
increase the scan
time and will lower
the performance.
Select a Excitation
wavelength in the
low UV
(230...260 nm). This
will cover nearly all
fluorescence in your
sample.