EasyManua.ls Logo

Agilent Technologies 1100 Series - Page 70

Agilent Technologies 1100 Series
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
72 1100 Series FD Reference Manual
3 Optimizing the Fluorescence Detector
Because of a known set of applications a 295 nm cut-off filter is built-in for
undisturbed application up to 560 nm without compromises (see Figure 31 on
page 72).
Figure 31 Reducing Stray Light
Stray light
Exitation 300nm
Second order
light 600nm
no filter
filter 280 nm
LU
Wavelength [nm]
F
lu
o
re
s
c
e
n
c
e

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Related product manuals