EasyManua.ls Logo

Agilent Technologies 8890 - Page 265

Agilent Technologies 8890
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
17 Chromatographic Checkout
Operation Manual 265
3 Ignite the FPD+ flame, if not lit.
4 Display the signal output and monitor. This output typically runs around 10. Wait for the
output to stabilize. This takes approximately 1 hour.
If the baseline output is too high:
Check column installation. If installed too high, the stationery phase burns in the flame
and increases measured output.
Check for leaks.
Bake out the detector and column at 250 °C.
Wrong flows set for installed filter.
If the baseline output is zero, verify the electrometer is on and the flame is lit.
5 If using a data system, prepare the data system to perform one run using the loaded
checkout method. Make sure that the data system will output a chromatogram.
ALS settings (if installed)
Sample washes 2
Sample pumps 6
Sample wash volume 8 (maximum)
Injection volume 1 µL
Syringe size 10 µL
Solvent A pre washes 2
Solvent A post washes 2
Solvent A wash volume 8
Solvent B pre washes 0
Solvent B post washes 0
Solvent B wash volume 0
Injection mode (7693A) Normal
Airgap Volume (7693A) 0.20
Viscosity delay 0
Inject Dispense Speed (7693A) 6000
PreInjection dwell 0
PostInjection dwell 0
Manual injection
Injection volume 1 µL
Data System
Data rate 5 Hz
Table 47 FPD+ Phosphorus Checkout Conditions (continued)

Table of Contents

Other manuals for Agilent Technologies 8890

Related product manuals