174 Chapter5
Language Reference
Common IEEE Commands
Query Instrument Options
*OPT?
Returns a string of all the installed instrument options.It is a comma
separated list such as: BAC,BAH. There are a few options that include
more then one mode. An istrument with one of these options will report
the option number once for each mode. You would get a response:
BAC,BAE,BAE,BAH For an instrument that contains cdmaOne (BAC),
NADC (BAE), PDC (BAE), and GSM (BAH).
Recall
*RCL <register>
This command recalls the instrument state from the specified
instrument memory register.
Range: registers are an integer, 0 to 19
Front Panel
Access:
File, Recall State
Reset
*RST
This command presets the instrument to a factory defined condition
that is appropriate for remote programming operation. *RST is
equivalent to performing the two commands :SYSTem:PRESet and
*CLS.
The :SYSTem:PRESet command is equivalent to a front panel
Preset.
The front panel
Preset sets instrument parameters to values for good
local/human interaction. The *RST and front panel
Preset will be
different. For example, the *RST will place the instrument in single
sweep while the front panel
Preset will place the instrument in
continuous sweep.
Front Panel
Access:
Preset

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals