Chapter 5 335
Language Reference
SENSe Subsystem
Spectrum—Frequency Span
[:SENSe]:SPECtrum:FREQuency:SPAN <freq>
[:SENSe]:SPECtrum:FREQuency:SPAN?
Set the frequency span to be measured.
Factory Preset
and *RST: 1.0 MHz
100.0 kHz for iDEN mode
Range: 10 Hz to 10.0 MHz (15 MHz when Service mode is
selected)
Default Unit: Hz
Remarks: The actual measured span will generally be slightly
wider due to the finite resolution of the FFT.
To use this command, the appropriate mode should be
selected with INSTrument:SELect.
Spectrum—Sweep (Acquisition) Time
[:SENSe]:SPECtrum:SWEep:TIME <time>
[:SENSe]:SPECtrum:SWEep:TIME?
Set the sweep (measurement acquisition) time. It is used to specify the
length of the time capture record. If the specified value is less than the
capture time required for the specified span and resolution bandwidth,
the value is ignored. The value is set at its auto value when auto is
selected. This is an advanced control that normally does not need to be
changed.
Factory Preset
and *RST: 188.0 µs
15.059 ms, for iDEN mode
Range: 100 ns to 10 s
Default Unit: seconds
Remarks: NOTE: You must be in the Service mode to use this
command. Use INSTrument:SELect to set the mode.
This command only effects the RF envelope trace.

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals