348 Chapter5
Language Reference
SERVice Subsystem
Unlock Calibration Files
:SERVice[:PRODuction]:CALibrate:END
Unlocks all of the calibration files.
Set Default Calibration Data
:SERVice[:PRODuction]:CALibrate:INITialize <cal_fid>
Loads the specified calibration data with default values.
Range: cal_fid, corresponds to the calibration data file ID
Remarks: No query.
Store Calibration Data in EEROM
:SERVice[:PRODuction]:CALibrate:STORe <cal_fid>
Stores the specified calibration data into EEROM. The data will survive
power cycles and will be reloaded into NRAM on power up.
Range: cal_fid, corresponds to the calibration data file ID
Remarks: No query.