202 Chapter5
Language Reference
CALibration Subsystem
Auto Adjust the Internal 10 MHz Frequency
Reference
:CALibration:FREQuency:REFerence:AADJust
Auto adjustment of the internal frequency reference (10 MHz
timebase). .
Remarks: You must be in the Service mode to use this command.
Use INSTrument:SELect.
Requires the current measurement to be timebase
frequency. A valid password needs to be entered
sometime prior to sending this command. See the
timebase frequency measurement for more
information.
Front Panel
Access: Select
Timebase Freq under Measure, then press Meas
Setup, Auto Adjust Now.
Align the ADC
:CALibration:GADC
:CALibration:GADC?
Performs the ADC group of alignments. The query returns a 0 if the
alignments occurred without problems.
Front Panel
Access:
System, Alignments, Align Subsystem, Align ADC
Align the IF Gain
:CALibration:GAIN:IF
:CALibration:GAIN:IF?
Calculate the curve coefficients for the IF gain DAC.
Front Panel
Access:
System, Alignments, Align Subsystem, IF

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals